MasterCraft 200V Manuale dell'Operatore Pagina 4

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 17
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 3
MY-60 MY-60T MY-61 MY-62 MY-63 MY-64
200
2K
20K
200K
2M
20M
200M
1.0%±5
5.0%±20
Range
Accuracy
1.0%±10
1.0%±3
7) TEMPERATURE (MY-60T/MY-62/MY-64 only)
MY-60T MY-62 MY-64
-20
~ 400
400
~ 1000
Range
Accuracy
0.75%±3
1.5%±15
With K-type thermocouple wire
8) FREQUENCY TEST(MY-63 & MY-64 only)
MY-63 MY-64
2KHz
20KHz
Range
Accuracy
1.5%±10
Sensitivity: 100mV rms
4. PRECAUTIONS AND PREPARATIONS FOR MEASUREMENT
1) Be sure that battery is correctly placed in the battery case and
connected to the battery snap.
2) Don’t exceed the input limit shown below:
Function Range Input terminals Maximum input
DCV 200mV 250VDC
ACV 200mV 250VAC
DCV 2~1000V 1000VDC
ACV 2~700V 700VAC
OHM V/OHM COM
Freq. V/OHM/Hz COM
Diode V/OHM COM
DCA 200mA
ACA 200mA
DCA 2A
ACA 2A
DCA10A
ACA10A
10A DC/AC10A COM
V/OHM COM
250V DC/AC
A COM
200mA DC/AC
2A DC/AC
3) Inspect the test leads for damaged insulation or exposed metal. Check
Test lead continuity. Damaged leads should be replaced.
4) Select the proper function and range for your measurement.
5) Check the input terminal position for red test lead depends on